BLOOD TESTING AT THE SINGLE CELL LEVEL USING QUANTITATIVE PHASE AND AMPLITUDE MICROSCOPY, BIOMED. OPT. EXP, 2(12), 2011

M. MIR, K. TANGELLA AND G. POPESCU
2011

 

It has recently been shown that quantitative phase imaging methods can provide clinically relevant parameters for red blood cell analysis with unprecedented detail and sensitivity. Since the quantitative phase information is dependent on both the thickness and refractive index, a major limitation to clinical translation has been a simple and practical approach to measure both simultaneously. Here we demonstrate both theoretically and experimentally that, by combining quantitative phase with a single absorption measurement, it is possible to measure both quantities at the single cell level. We validate this approach by comparing our results to those acquired using a clinical blood analyzer. This approach to decouple the thickness and refractive index for red blood cells may be used with any quantitative phase imaging method that can operate in tandem with bright field microscopy at the Soret-band wavelength.

Scroll to top
Phi Optics
Privacy Overview

This website uses cookies so that we can provide you with the best user experience possible. Cookie information is stored in your browser and performs functions such as recognising you when you return to our website and helping our team to understand which sections of the website you find most interesting and useful.